Quick Reference: Metric Decision Guide - ScreenAssist
Use the table below to quickly identify which metric to interrogate when a specific quality issue in ScreenAssist is observed.
If you observe… | Check this metric first: | Likely cause: |
Low Z’ but high Robust Z’ | Classical vs. Robust Z’ comparison | Outlier wells in control population |
High Neutral RSDE | Fluorescence CV of reference wells | Inconsistent dye loading or sample variability |
Low S/N on titration curve | Response Amplitude (Δ Ratio) | Insufficient assay window; revisit target/dye concentration |
High Δ Fluorescence (titration or single point) | 650 nm fluorescence across all wells | Compound autofluorescence, quenching, or aggregation |
Many poor reads | Well scan quality scores | Capillary blockage, precipitation, or instrument issue |
|SSMD| < 6 but Z’ ≥ 0.5 | Control population variances | Unequal variances between positive control and reference |
Marginal Z’ on multiple sequential plates | Plate-by-plate Z’ trend over time | Assay signal drift — check protein or dye stability over screen duration |